en Language
Home > News > How to Reduce Air Consumption in Pneumatic Actuator Valves

How to Reduce Air Consumption in Pneumatic Actuator Valves

2025-04-08

Here are the steps to calibrate a pneumatic actuator valve for optimal performance:

Preparation: Ensure the valve is installed correctly and the pneumatic actuator is connected properly. Check the air supply system to make sure it provides stable pressure and clean air.

Zero - point calibration: With the valve in the closed position, adjust the zero - point of the positioner or controller. This is usually done by setting the output signal of the controller to the minimum value that corresponds to the closed - valve position.

Full - scale calibration: Supply the maximum air pressure to the actuator to open the valve fully. Then, adjust the full - scale setting of the positioner or controller so that the valve reaches its maximum opening position as indicated by the process requirements.

Intermediate - point calibration: Check and adjust the valve position at several intermediate points between the fully closed and fully open positions. This ensures that the valve responds accurately and linearly to the control signals within its entire operating range.

Fine - tuning: Observe the valve's performance during actual operation and make any necessary fine - tuning adjustments to optimize its response time, accuracy, and stability.

Documentation: Record all the calibration settings and results for future reference and maintenance.

If you want to learn more about low-priced products, please visit the following website: www.xm-valveactuator.com

Send Inquiry
code